Sentinel C28WE

Trademark: CTS
Origin: USA


2611 lượt xem

The Sentinel C28WE pressure decay or vacuum decay instrument is Cincinnati Test System’s latest addition to the technologically advanced Sentinel leak test instrument family. The Sentinel C28WE utilizes test algorithms processed on an advanced 32-bit system, giving it fast and repeatable results in a more compact package.

Testing Technologies of The Sentinel C28WE

Single Test Configurable for Pressure or Vacuum Decay Test Requirements                

  • Pressure/Vacuum decay-pressure change
  • Pressure/Vacuum decay-with certified mass flow leak standard
  • Pressure/Vacuum decay-pressure change/time
  • Occlusion test (vacuum or pressure test)
  • Storage for 32 pre-programmed test sequences
  • Test Program can be selected through keypad, or binary input pulse
  • Test results evaluation is fixed (Fail/Pass/Fail)
  • Test result output format is configurable

Selectable Test Parameters

  •  Quick Test
  •  Test Pressure Level Compensation
  • Adjustable Environmental Drift Correction per test program

Hystrong is the Only Authorized Unit of Cincinnati Test Systems (CTS) in Vietnam, with a team of experienced Technicians.
Contact Hystrong Vietnam for advice on installing the most suitable system! Contact now>>>

Facebook: HystrongICL

Hotline: (+84) 919-995-040 (call/zalo)

Features Measurement Range Connectivity
Advanced Product Features

  • Wall mount configuration 9” W x 6.5” H x 7.25” D
  • Powerful 32-bit processor & 24 –bit A/D converter
  • Language neutral color LCD 12 line 480 x 272 operator display
  • Bar Graph display  (pressure, loss, or flow)-English, Chinese, Korean, Spanish
  • Individual LED Pass/Fail/In-test test status indicators
  • Expanded instrument control through EtherNet/IP™ Communication
  • Modular pneumatic and control packages
  • Single test channel

Standard Features

  • Auto Calibration includes mass flow leak standard certified for pressure and flow
  • Transducer zero and span are automatically reset during auto calibration
  • “Performance factor” calibration feedback ensures accurate results
  • Manual edit of Calibration Program permitted
  • Self Test isolates the test instrument pneumatics to verify a “no leak” condition
  • Features ensure accurate results and easy operator setup and troubleshooting
  • Firmware validation document included
Standard Test Pressure Ranges

Standard pressure transducer, valve manifolds, and regulator ranges: -14.5, 2, 30, 100, 200, or electronic regulator (PSIG)

  • Optional High cV flow valves and manifold available up to 100 psig range allows 5 times faster fill and exhaust to reduce cycle times on large parts
  • Absolute pressure transducer
  • Two (2) point pressure calibration
  • Data resolution 0.00001 psi/ and scc/m.
  • Optional electronic regulator available with multi-point calibration

Selectable Test Units of Measure

  • Pressure: PSIG, PSIV, inH2O, cmH2O, mmH2O, kg/cm2, ATM, inHg, cmHg, mmHg, kPa, Pa, MPa, Bar, mBar, Torr
  • Flow: Sccm, sccs, scch, slpm, slps, slph, scfm, scfs, scfh
  • Time: Sec, msec, min
  • Units of measure can be instrument global or per test program
Tooling Control

  • Expanded control through EtherNet/IP™ Communication
  • 6 inputs/3 outputs – 1 programmable tooling motion – no feedback
  • Tooling I/O is  internally connected
  • I/O 24 VDC – optional I/O cable “Cable,CTS, P56-540”

Test Data Storage

  • Up to 5000 LIFO part results including Date / Time / Test Results
  • SPC data tracking for part traceability
  • External test data storage and wave form analysis through interface with the QualityWorX CTS DataHub

Electrical Requirements and Connectivity

  • 88-264 VAC, 47-63 Hz or optional 22-26 VDC
  • RS232(1)
  • Telnet(1)
  • USB storage device for firmware upload only (FAT32, USB1.0,1.1,2.0)
  • ¼” Colder fitting for test line attachment

Reviews

There are no reviews yet.

Be the first to review “Sentinel C28WE”

    REQUEST A QUOTE