Sentinel Blackbelt

Trademark: CTS Việt Nam
Origin: USA


2593 lượt xem

The CTS Blackbelt is a precision benchtop leak test instrument designed for maximum performance and minimal space requirements. Cincinnati Test Systems’ Blackbelt instrument utilizes advanced algorithms derived from over 30 years of proven leak test experience to produce the fast and accurate results required by the most demanding customers.

Testing Technologies

Multi-Test Configurable for Pressure Decay, Vacuum Decay or Mass Flow Test Requirements

  • Pressure/Vacuum decay-pressure change
  • Pressure/Vacuum decay-with leak standard
  • Pressure/Vacuum decay-pressure change/time
  • Pressure decay-differential pressure loss
  • Occlusion test (vacuum or pressure)
  • Mass flow
  • Mass flow continuous direct flow reading
  • Mass flow with certified mass flow leak standard
  • Ramp up or down to change in pressure event
  • Ramp to flow event with direct flow measurement
  • Ramp to “Proof” test
  • Sequential Test Linking
  • “Parent Program” holds a customized test sequence for your multi-test application
  • Storage for 99 pre-programmed test sequences
  • Test Program can be selected through keypad, RS232, or Binary
  • Test results evaluation is Selectable (Fail/Pass/Fail etc)
  • Test result output format is configurable format

Selectable Test Parameters

  • Quick Test
  • Test Pressure Level Compensation (selectable on/off)
  • Adjustable Environmental Drift Correction per test program

—-

Hystrong is the Only Authorized Unit of Cincinnati Test Systems (CTS) in Vietnam, with a team of experienced Technicians.
Contact Hystrong Vietnam for advice on installing the most suitable system! Contact now>>>

Facebook: HystrongICL

Hotline: (+84) 919-995-040 (call/zalo)

Features Measurement Range Connectivity

Advanced Features

  • Benchtop configuration 9” W x 11.25” H x 15” D
  • Powerful 32-bit processor & 24 –bit A/D converter
  • Language neutral color LCD 12 line 480 x 272 operator display
  • XY Graphing (pressure, loss, or flow)-English, Chinese, Korean, Spanish
  • Light ring indicators per test port identifies test port status Pass/Fail/In-test
  • Modular pneumatic and control packages
  • Expanded instrument control through EtherNet/IP™ Communication
  • Single test channel, with multiple test port options(maximum of 4)
  • Barcode ready input RS232 Alpha-numeric string stored with test result.

Standard Features

  • Auto setup, verifies test pressure and selects the best test timers for optimum results in the allowed time cycle (US Patent #7,594,424 B2)
  • Auto Calibration includes mass flow leak standard certified for pressure and flow
  • Transducer zero and span are automatically reset during auto calibration
  • Optional “External leak Standard” port 
  • “Performance factor” calibration feedback ensures accurate results
  • Manual edit of Calibration Program permitted
  • Self Test isolates the test instrument pneumatics to verify a “no leak” condition
  • Features ensure accurate results and easy operator setup and troubleshooting
  • Firmware validation document included

Standard Test Pressure Ranges

Standard pressure transducer, valve manifolds, and regulator ranges:
-14.5, -2, 0.5, 2, 10, 30, 100, 200, 400, 500 or electronic regulation (PSIG)

  • Optional low volume test manifold available
  • Absolute pressure transducer with accuracy of +/-0.05% of full scale
  • Six point pressure transducer calibration
  • Data resolution 0.00001 psi/ and scc/m.
  • Optional electronic regulator available with multi-point calibration
  • Optional internal vacuum generator available
  • Optional automatic supply shutoff available

Mass Flow Sensor Ranges

– CTS will optimize the flow sensor & flow range per application. Contact Application Engineering for available mass flow sensors.

Pressure Selections

PSIG, PSIV, inH2O, cmH2O, mmH2O, kg/cm2, ATM, inHg, cmHg, mmHg, kPa, Pa, MPa, Bar, mBar, Torr

Flow Selections

Sccm, sccs, scch, slpm, slps, slph, scfm, scfs, scfh

Time Selections

Sec, msec, min

Units of measure can be instrument global or per test program

Tooling Control

  • Expanded control through EtherNet/IP™ Communication
  • 6 inputs/6 outputs with up to five programmable tooling motions with feedback
  • Optional “CTS Connect” pneumatic actuator port (5 pneumatic outputs maximum)
  • Tooling I/O is  externally connected
  • I/O 24 VDC – optional I/O cable “Cable,CTS, P56-540”

Test Data Storage

  • Up to 5000 LIFO part results including Date / Time / Barcode / Test Results
  • Selectable part serialization in the test data stream
  • SPC data tracking for part traceability
  • Expandable download through USB port
  • External test data storage and wave form analysis through interface with the QualityWorX CTS DataHub
  • Key lock or password protected program and data security

Electrical Requirements and Connectivity

  • 88-264 VAC, 47-63 Hz or optional 22-26 VDC
  • RS232(1)
  • Telnet(1)
  • Email Alerts
  • Barcode ready (RS232)
  • Mobile text
  • USB mass storage device for backup, restore, and reports (FAT32, USB1.0,1.1,2.0)
  • 1/8” NPT, Male Luer or Female Luer fitting for test line attachment