INFICON XL3000FLEX

Trademark:
Origin:


333 lượt xem

INFICON XL3000FLEX – Helium and Hydrogen Leak Detector

INFICON XL3000FLEX integrates advanced HIGH FLOW Technology. The device provides the highest Accuracy measurement results with the fastest Speed!

The XL3000FLEX leak detector was the first to detect leaks with gas formation (95% nitrogen and 5% hydrogen) as accurately as with helium. with the smallest detectable leak rate of 1 ∙ 10-7 mbar∙l/s for both gases.

XL3000flex using the High Flow of INFICON help device is active with save volume 3000 sccm. It is the highest current on-field air flow that can be tested. This is allow to move the start up on the sample try at speed of faster and range of Farther. So even the most inexperienced operators can achieve accurate results with the XL3000flex.

An additional advantage of testing high flow gas flows is that robotic leak detection/checking can be enabled.

Advantage of INFICON XL3000FLEX 

  • Integrated HIGH FLOW INFICON . Technology
    • Provides the most accurate detection results – Avoid missing leaks!
    • Detects leaks even in hard-to-reach areas.
    • Supports even inexperienced users who can reliably locate leaks.
    • Ideal for robotic inspection operations.
  • Minimize your operating costs – choose the tracer gas of your choice
  • Easy integration with Robot operation with Fieldbus

HystrongICL is a leading distributor of INFICON instruments in Vietnam, with an experienced technical team

Contact HystrongICL for more information! Contact now>>>

Facebook: HystrongICL

Hotline: (+84) 865.733.288 (call/zalo)

Specifications 

TYPE   XL3000flex
MINIMUM DETECTABLE LEAK RAT mbar•L/s 2E-7 (low flow)
MINIMUM DETECTABLE LEAK RATE mbar•L/s 2E-6 (high flow)
GAS FLOW sccm 300 (low flow) – 3000 (high flow)
ION SOURCE   2 long life yttria-coated iridum filaments
MEASURABLE GASES   Helium, H2
OPERATING VOLTAGE   100-120V (ac) 50/60 Hz 230V (ac) 50/60 Hz
POWER VA 280
RESPONSE TIME   <1 s
RUN UP TIME   150 s

 

Product Code & Components

PART NUMBER Description
12339 External Sniffer Test Leak H2/HeCal leak
520-200 Sniffer Leak detector XL3000 flex
520-250 CalMate
520-250 CalMate
521-015 Adapter customer sniffer line P3000XL
521-018 Sniffer tip ST312XL, 120 mm, rigid
521-019 Sniffer tip FT312XL, 120 mm, flexible
521-020 Sniffer tip ST385XL, 385 mm, rigid